Why does transit time insertion sensors adopt Z method instead of V method?

There are four ways to install transit time ultrasonic sensors, V method and Z method are commonly used, and Z method is used to install transit time insertion sensors on site. This is mainly due to the insertion type sensors installation characteristics and Z method signal transmission mode. When installing the inline sensors, it is necessary to weld the ball valve base, install the ball valve, and cut holes in the pipeline. Therefore, once the installation position is determined, it cannot be the same as the external clamp on type. When the signal is bad, the sensor installation mode can be changed, that is, the sensor installation position can be changed. Therefore, when installing the insertion sensors, choose the installation mode with the strongest signal, namely the Z method.